Other solvents

Other solvents
Topic LCI-USG-0115, Last Updated 05/19/2025

This list comprises solvents you can use in your system after installing a Waters compatibility kit.

Note: Without the appropriate compatibility kit, these solvents can shorten equipment life. If you routinely use the solvents on this list, Waters recommends that you install the Waters hexane/tetrahydrofuran (THF) compatibility kit applicable to your system.
  • Acetone
  • Ethyl acetate
  • Hexane
  • THF

When you change solvents, consider solvent polarity. When switching between polar and nonpolar solvents, flush the system with a miscible and compatible solvent, such as IPA.


Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Refer to the English version of this page for the latest information.