Installing the (optional) metal column outlet tube

Installing the (optional) metal column outlet tube
Topic LCI-USG-0223, Last Updated 11/21/2025

Users or Waters field service personnel can replace the standard PEEK column outlet tube with an optional metal column outlet tube.

Required tools and materials

  • Metal column outlet tube (see Optional replacement column outlet tubes for the type of metal compatible with your system)
  • Chemical-resistant, powder-free gloves
  • 1/4-inch open-end wrench
  • T10 TORX driver

To install the metal column outlet tube in the column compartment:

Warning: To prevent burn injuries, allow sufficient time for the column to cool before opening the compartment door. The column, compartment, tubing, fittings, and door liner can be hot.
Warning: To avoid personal contamination with biologically hazardous, toxic, or corrosive materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
  1. Stop primary solvent flow through the system and allow the column and compartment temperature to drop below 55 °C.
  2. Open the column compartment door.
  3. Disconnect the column outlet tube's fitting from the detector's flow cell inlet.
  4. Disconnect the tube from the column, and then remove it from the two clips in the detector and the four fasteners at the top of the column compartment.
  5. Install the replacement metal column outlet tube by reversing the actions performed in the previous two steps:
    1. Connect the tube to the column, and then route it into the column compartment's top four fasteners and the detector's two clips.
      Figure 1: Installed metal column outlet tube routed through compartment's top fasteners
      Column compartment's top fasteners (circled in red)
    2. Reconnect the new tube's tool-free fitting (TFF) to the detector's flow cell inlet.
  6. Close the column compartment door.

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Refer to the English version of this page for the latest information.