Shutting down for more than 24 hours

Shutting down for more than 24 hours
Topic LCI-USG-0214, Last Updated 11/21/2025

Failure to flush the system/detector before shutdown can result in a plugged flow cell.

To shut down the detector for more than 24 hours:

  1. Extinguish the detector lamp by tapping COMMANDS > UV Detector Lamp.
    Power off Lamp appears on the screen.
  2. Remove buffer salts and additives by flushing with water.
  3. Flush the column and flow cell with 100% pure organic solvent.