Priming the pump is part of the system startup workflow on the touchscreen or the console.
Priming prepares a new system for use or for a change in reservoirs or solvents. It also prepares a system for restarting after it is idle for more than four hours. During priming, the vent valve moves to the vent position, allowing flow to go to waste. The flow rate during priming is 10 mL/min. See Running the startup workflow.
Ensure that the solvent reservoirs contain sufficient solvent for adequate priming and subsequent operation of the system, and that the waste container can hold all the used solvent. For example, at 10 mL/min, priming for two minutes uses about 20 mL of each solvent.