Choosing extension loops

Choosing extension loops
Topic LCI-USG-0145, Last Updated 11/21/2025

An extension loop affects injection volume and system pressure. Be sure to select the correct loop for your application.

Extension loops, which increase the volume of sample that can be drawn and held for injection, are an optional part of the injection system. You install them between the needle and the pressure transducer.
Notes:
  • On systems with the multi-draw kit, the extension loops are installed between the needle and the multi-draw valve.
  • Any injection greater than 100 µL requires the multi-draw valve option.
Table 1. The following extension loops are available to use in the sample manager
Loop size 1 2
50 µL
100 µL - Standard 3
250 µL
1000 µL
2000 µL