Waters offers a comprehensive range of analytical system solutions, software, and services for scientists. Liquid Chromatography. Mass Spectrometry. Waters is the leading provider of lab equipment, supplies and software for scientists across the world. Easily research and order everything your lab needs! Waters offers a comprehensive range of analytical system solutions, software, and services for scientists. Liquid Chromatography. Mass Spectrometry. Waters is the leading provider of lab equipment, supplies and software for scientists across the world. Easily research and order everything your lab needs!
Warning: To avoid eye injury, use eye protection when performing this procedure.
Note:
To avoid contaminating the flow cell, wear clean, chemical-resistant, powder-free gloves when handling, removing, or replacing it.
To avoid damaging the flow cell, handle it with care. Do not disassemble the flow cell.
Via the system's touchscreen, ensure that the detector lamp's power status is "off".
Tap MAINTAIN > Replace Components > Replace Flow Cell > NEXT > NEXT > START and follow the on-screen prompts.
Open the PDA detector's door, gently pulling its right-hand edge toward you.
Before putting it into storage, Waters recommends that you flush the old flow cell. Select a solvent compatible with the samples and mobile phases that you have been using. If you have been using buffers, flush the cell with 10 mL of HPLC-grade water, and then flush with 10 mL of a low-surface-tension solvent, such as methanol.
Requirement: Ensure that the solvent used is miscible with the previous mobile phase.
Disconnect the PDA detector’s inlet and outlet tubing.
Figure 1: Locating the flow cell (PDA detector door open view)
Outlet tubing
Flow cell assembly
Inlet tubing
Flow cell handle
Thumbscrews (3)
Remove the flow cell:
Using the screwdriver, loosen the three thumbscrews on the flow cell assembly’s front plate.
Remove the flow cell by grasping its handle and gently pulling the assembly toward you.
Unpack and inspect the new flow cell, ensuring that the flow cell type is correct for your application.
Note: When replacing the flow cell, replace the flow cell inlet tubing with the tubing included with the new flow cell.
Install the new flow cell:
Align the flow cell assembly with the front of the opening, and then insert it slowly so that its two alignment receptacles (on the rear face of the front part of the cell flange) engage the pins in the cell compartment.
Tip: As you insert the cell, its drip tray helps ensure proper alignment of the compartment's pins and the cell flange's receptacles.
Figure 2: Installing the PDA detector's flow cell assembly
Thumbscrew (3)
Compartment alignment pin (2)
Continue inserting the flow cell until the three thumbscrews align with their holes in the bulkhead.
Figure 3: Aligning the detector's flow cell assembly
Cell flange alignment receptacle (2)
Thumbscrew (3)
Hand tighten the thumbscrews, and then verify that the screws are secure using the screwdriver.
Connect the inlet tubing to the main column connection and flow cell inlet, and then connect the outlet tubing to the flow cell outlet.
Ensure that the flow cell is filled with degassed, transparent solvent (acetonitrile or water) that is free of air bubbles.
Note: If verification fails, troubleshoot and re-attempt verification. If verification still fails, perform an erbium calibration. See
Erbium calibration.
To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.