Installing the tool-free fitting adapter

Installing the tool-free fitting adapter
Topic LCI-USG-0231, Last Updated 11/21/2025

Waters BioSuite columns and some non-Waters columns may leak when connected with the tool-free fitting. In this case, the user must install an adapter between the column and the tool-free fitting.

Warning: To avoid personal contamination, wear clean, chemical-resistant, powder-free gloves when handling fittings.

Required tools and materials

  • Appropriately sized open-end wrench to counterhold column
  • 5/16-inch open-end wrench
  • Tool-free fitting adapter (separate adapters are available for Alliance iS and Alliance iS Bio)
Note: The tool-free fitting adapter is a single-use device. A new adapter is required for each additional column.

To install the tool-free fitting adapter:

  1. Push the end of the adapter inward until it is flush with the O-ring.
    Figure 1: Tool-free fitting to column adapter
    End of tool-free fitting adapter
    O-ring
  2. Tighten the tool-free fitting all the way into the adapter.
    Figure 2: Tightening the tool-free fitting to the adapter
  3. Unscrew the tool-free fitting from the adapter by 1-1/2 turns.
    Figure 3: Unscrewing the tool-free fitting from the adapter
  4. Push the end of the adapter inward until it stops against the tool-free fitting.
    Figure 4: Pushing the end of the adapter in
  5. Unscrew and remove the tool-free fitting from the adapter.
    Figure 5: Removing the tool-free fitting from the adapter
  6. Remove the O-ring from the adapter.
    Caution: To avoid leaks, ensure that the tube remains stationary when removing the O-ring from the adapter. If the tube moves, reinstall the O-ring and repeat the procedure.
  7. Screw the adapter finger-tight into the column inlet port, and then counterhold the column inlet port with the appropriate size wrench. Using the 5/16-inch open-end wrench, tighten the tool-free fitting adapter an additional 3/4-turn.
    Figure 6: Tightening the tool-free fitting adapter to the column inlet port
    Column fitting
    Tool-free fitting adapter
  8. Tighten the tool-free fitting securely into the adapter. For more information on tool-free fittings, see Tool-free fitting.
    Figure 7: Installing the tool-free fitting to the adapter

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Refer to the English version of this page for the latest information.