Calibrating the needle z axis

Calibrating the needle z axis
Topic LCI-USG-0181, Last Updated 11/21/2025

Users or Waters field service personnel can calibrate the sample manager needle's z axis.

You must calibrate the needle before you use the sample manager for the first time and whenever you replace the sample needle. Failing to calibrate the needle can damage it. The calibration procedure is identical for all needles.

To calibrate the needle z axis, do the following:

  1. From the touchscreen, tap Maintain > Calibrate needle Z axis.
  2. Follow the directions on the screen.

Need Help? Connect with a Waters expert. 