TUV detector optics

TUV detector optics
Topic LCI-USG-0112, Last Updated 05/19/2025

The detector optics are based on a Fastie-Ebert monochromator.

The following figure shows the components that make up the detector optics assembly.

Note: In addition to what is shown in the figure, detector optics also include a shutter, wavelength calibration filter, and second-order filter.
Figure 1: Detector optics assembly
Mirror
High-brightness deuterium (D2) lamp
Window
Filter wheel
Slit
Mirror
Grating
Reference photodiode
Beam splitter
TaperSlit flow cell
Sample photodiode