2D and 3D acquisition

2D and 3D acquisition
Topic LCI-USG-0218, Last Updated 05/19/2025

The Alliance iS system PDA detector is capable of both 2D and 3D acquisition.

The PDA detector can simultaneously collect two types of data channels: spectra (3D) and chromatograms (2D). For best results in library matching and peak-purity analysis, set the 3D resolution to 1 nm.

For chromatograms (2D data), select a resolution to optimize signal amplitude, baseline noise, and linear dynamic range. When the monitoring wavelength for an analyte corresponds to a peak’s lambda max, increasing bandwidth tends to decrease peak height while reducing baseline noise and linear dynamic range.

Tip: A resolution of 4 nm is effective for many analytes.
Figure 1: Resolution comparison for caffeine

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Refer to the English version of this page for the latest information.

Back To Top Back To Top