Variable slit width

Variable slit width
Topic LCI-USG-0212, Last Updated 05/19/2025

Variable slit width allows the user to balance resolution and throughput to best fit their analysis.

This Alliance iS PDA detector features an actuator-controlled variable slit. The slit width sizes are:

  • 35 µm
  • 50 µm (default)
  • 100 µm
  • 150 µm
Figure 1: Alliance iS PDA variable slits

A small slit provides the sharpest spectral peak resolution while a large slit provides the best signal with respect to noise for improved sensitivity.

Figure 2: Impact of slit width on resolution of benzene

The slits are designed to deliver energy levels proportional to their areas.